FF-1513-3 usa jmk inc 15 caldwell dr. amherst, nh 03031 p h: 603 886-4100 fx: 603 886-4115 email: info@jmkfilters.com eur jmk inc glasgow g13 1dn scotland uk ph: 44-(0) 7785310729 fax: 44-(0) 141 589 1884 operating voltage (ac)= ______________ 115 / 250 v~ operating current= ______________________ 3.0 amp operating frequency= ___________________ 50/60 hz operating temperature= _______________-25 to 50c diel. withstanding (line - case)= ___________ 1500 vac diel. withstanding (line - line)= ____________ 1500 vdc leakage current= _____ 60 ua max. @ 125 vac, 60 hz = ____ 100 ua max. @ 250 vac, 50hz max residual voltage after 1 sec= __________ 34 vol ts calcultaed mtbf= ________ 1,306,174 hrs @ 30c gb the FF-1513-3 filter has been designed to operate a t the input power point of medical equipment. the filter features low leakage current and the bas ic filter is configured to be compatible with fcc and en55022 requirements. agency approvals= ul order p/n: FF-1513-3 insertion loss vs. frequency brn blue brn blue load line grn/y brown brown blue g/y g/y blue awg #18 (5pl) blue 12.0 0.5 12.0 0.5 1.75 3.00 2.19 3.43 3.80 ?0.200 (2x) jmk, inc. FF-1513-3 125/250 vac 3 amp 50c line load assy in mexico line load
|